The response of a position-sensitive Li-glass scintillator detector to alpha-particles from a collimated 241Am source scanned across the face of the detector has been measured. Scintillation light was read out by an 8x8 pixel multi-anode photomultiplier and the signal amplitude for each pixel has been recorded for every position on a scan. The pixel signal is strongly dependent on position and in general several pixels will register a signal (a hit) above a given threshold.

 